Weidmuller PTX800A - Configurable counter and totalizer Manual de usuario Pagina 42

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 48
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 41
42
Procedure
When the
display shows
Action/Description
Put the instrument in setup mode and scroll through the main menu
CALIn_n
Press ì or Å
CAlIn_y
Press PGM to select Input calibration
InP_LO_=
Connect the multimeter to measure the Input level, then press PGM
4.00
Set the Input signal to the appropriate value, e.g., 4.00 mA
When you are happy that the Input is correct, press PGM
Inp_HI_=
After approx 5s the display introduces the next calibration point. Press PGM
20.00
Set the Input signal to the appropriate value, e.g., 20.00 mA
When you are happy that the Input is correct, press PGM
dFactor=
Continue with the setup sequence.
Note: Do not remove the power while the save message is on display.
PTX800 analogue output board showing output type jumper locations.
Link both pairs
of posts for
voltage outputs
Vista de pagina 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 48

Comentarios a estos manuales

Sin comentarios