37
• Installation
Caution: In order to meet product safety requirements, these units must
only be installed, by qualified staff, in accordance with the information
given in this manual, using the mounting clips and terminal blocks
supplied, and all relevant national electrical wiring and safety rules
must be followed.
Locate the instrument in an area that is free from dust, moisture
and corrosive gases (pollution degree II or better). Do not cover the
ventilation holes at the side of the case.
Po
wer Supply Requirements
Check the power supply against the model number before applying
power to the instrument. PTX800A displays must be used with a
suitable mains Power Supply, ratings are as follows:
Voltage Range Nominal Voltage Power Frequency
100–132 V AC 110 V AC 6 VA 47-63 Hz
200–264 V AC 240 V AC 6 VA 47-63 Hz
21.6–26.4 V DC* 24 V DC 6 W (at 24 V DC) -
*The 24 V DC rated model has cULus approval. File number E256486.
Digital pulse output
The retransmit pulse output gives a single pulse every time the total
in
creases by one unit. The pulse width is a constant 32 ms and the
minimum off time is 32 ms. The output is an isolated, NPN, open
collector transistor type and can switch a maximum of 50 V DC or a
current up to 200 mA.
Reset pulse input
Accepts a volt-free contact closure, or a shorting link, between pins
7 a
nd 12. The contact must be closed for at least 100 mS.