Weidmuller Portacal 275 - Signal source and loop calibrator Manual de usuario Pagina 13

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 32
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 12
13
4-20 mA range adjustment
1. Set the ‘current range select’ switch to 4-20 mA.
2. Set the ‘output level select’ switch to 0 %.
3. Set the ‘voltage/millivolt select’ switch to 200 mV.
4. Set the ‘source/sink mode’ switch to Source.
5. Adjust potentiometer A until the output is 4 ±0.001 mA.
6. Set the ‘output level select’ switch to 100 %.
7
. Adjust potentiometer B until the output is 20 ±0.001 mA.
8. Repeat the procedure until both readings are correct.
Voltage ranges
The adjustment of the current ranges will ensure that the voltage
ranges are correct.
Vista de pagina 12
1 2 ... 8 9 10 11 12 13 14 15 16 17 18 ... 31 32

Comentarios a estos manuales

Sin comentarios