40
When the
display shows
Action/Description
Put the instrument in setup mode and scroll through the main menu
C Pn
Press ì or Å
C Py
Press ENT to select input calibration
InLO
Set the signal source to the lowest value in the required range.
Wait a few seconds (for the signal to stabilise) and press ENT
InLO
Display will stop flashing (for about 20 s) to show that the instrument is processing
the input.
InHI
Set the signal source to the Highest value in the required range.
Wait a few seconds (for the signal to stabilise) and press ENT
InHI
Display will stop flashing (for about 10 s) to show that the instrument is processing
the input.
COPn
Continue with the setup sequence.
Note: To discard changes to calibration and setup you can remove the power at this point.
1. Gently move lug out a fraction with a screwdriver to release the
backplate.
2. Pull the backplate back slightly to keep lug from clicking back into
place.
3. R
epeat with each lug until backplate comes loose.
Analogue Input Calibration