Calibration procedure
When the
display shows
Action/Description ENT
ï
Put the instrument in setup mode
1.10
While the software version number is flashing,
remove the security link and press ENT
AnLg
Display indicates that unit is /AR type, press ENT
CIPn
Press Å or ì to Calibrate the inputs or ENT to skip
CIPy
Calibrate inputs? Start No
Connect the RTD Simulator to the inputs (for connections see over page).
OPEN
Leave inputs 2 & 3 open circuit Calibrate Skip
20.00
Connect 50 Ω resistor Calibrate Skip
5.00
Connect 200 Ω resistor Calibrate Skip
1.00
Connect 1 kΩ resistor Calibrate Skip
0.20
Connect 5 kΩ resistor Calibrate Skip
rCLy
Calibrate RTD Sensor? Start No
0C
Set the RTD Simulator to 0 °C Calibrate Skip
200C
Set the RTD Simulator to 200 °C Calibrate Skip
COPn
Press Å or ì to Calibrate the outputs or ENT to skip to the next setup item
When the
display shows
Action/Description
ì Å
ENT
Put the instrument in setup mode and scroll through the main menu
COPn
Press Å or ì
COPy
Calibrate Outputs? Toggle Start
Measure the output current
I=0
The output current should be 0.00 mA
(Use the keypad to adjust the value as required)
Increase Decrease Accept
I=20
The output current should be 20.00 mA Increase Decrease Accept
Measure the output voltage
E=0
The output voltage should be 0.00 volts Increase Decrease Accept
E=10
The output voltage should be 10.00 volts Increase Decrease Accept
rng=
Instrument will continue with the setup process
Note: Do not remove the power while the save message is on display.