• Operation
General
The CND/AR measures conductivity using conductance cell and Pt100
for temperature compensation. It displays the conductivity at the
reference temperature, produces a proportional analogue current/
voltage output and has two alarm channels.
Cleaning
The case can be wiped with a damp cloth. De-energise the unit before
cleaning.
Conductance (Ω
-1
), Conductivity (Ω
-1
cm
-1
) and Probe factors
The conductance measurement is the inverse of the resistance
between the two probes. This depends on properties of the probe, like
the distance between them. The probe factor P
f
gives the relationship
between measured conductance M
c
and the actual conductivity
measurement C, so that
C = M
c
x P
f
CND/AR block diagram
Cell Conductance
Analogue Output
Alarm 1
Alarm 2
Power Supply
Cell Temperature
(RTD)