Terminal Connections for output calibration
Analogue Output Channel Calibration Stage Signal type Terminal
Channel One
Current Output
mA output + 15
mA output − 16
Voltage Output
V output + 15
V output − 17
Channel Two
Current Output
mA output + 18
mA output − 19
Voltage Output
V output + 18
V output − 20
Output calibration procedure
1. Select the channel you wish to calibrate (see diagram above).
2. Press
ENT , display will show: A.Out1 @ 0.00 mA
3
. Measure the output current from that channel.
4. Use the
& and ^ keys to adjust the output until it is correct.
5. Press ENT , display will show: A.Out1 @ 20.00 mA
6.
Use the
& and ^ keys to adjust the output until it is correct.
7. Press ENT , display will show: A.Out1 @ 0.00 V
8
. Measure the output voltage from that channel.
9. Use the
& and ^ keys to adjust the output until it is correct.
10. Press ENT , display will show: A.Out1 @ 10.00 V
1
1. Use the
& and ^ keys to adjust the output until it is correct.
12. At the ‘Continue?’ prompt, select ‘Yes’ t
o go the next Input
Channel or ‘No‘ to go back to the calibration menu.
13. At the ‘Save?’ prompt, select ‘Yes’ t
o permanently save the
values or ‘No‘ to use the new values until next power up.