20
• Calibration
General
The URT/R is factory calibrated and you can select any input range
without recalibrating the unit. However, if you change the RTD type,
you must calibrate the inputs.
Allow the instrument 15 minutes of powered operation (to reach a
stable temperature) before calibration and, during calibration, allow a
few seconds for signals to stabilise before accepting the value.
Equipment requirements
• A suitable RTD simulator (accurate to 0.05 °C)
Input calibration points
Calibration point
Pt100 Ni120
Display Temperature (°C) Display Temperature (°C)
1
-200
-200.0 °C
-80n
-80.0 °C
2
700P
700.0 °C
320n
320.0 °C
3
-100
-100 °C
-80n
-80.0 °C
4
200P
200.0 °C
200n
200.0 °C
Terminal connections for input calibration
Calibration Stage Signal type Terminal
RTD Inputs
RTD Simulator A 2
RTD Simulator B 3
RTD Simulator Bsense 4